Centrifuge for Lab Automation

”Robo-Ready” Series for Lab Automation
Step Stop Centrifuges

In the future, lab automation is expected to increase.

Lab automation is expected to increase operational efficiency, reliability of results, and reproducibility.
However, the combination of lab automation and centrifuges is not a good relationship.
Although centrifuges are general-purpose laboratory instruments, they cannot stop the rotor at a defined position.
Step stop mixer-centrifuge stop the rotor at defined positions so that the robot arm can pick up the sample tubes.
TOMY-RESO will support the lab automation with Step Stop Centrifuges.
Robo-Ready is a series of centrifuges with a rotor stops at a defined position developed for lab automation.
This centrifuge is for lab automation.

Made In Japan

System Integrator
Friendly

定位置停止卓上遠心機 LCXR-100

Bench-Top Step Stop CentrifugeLCXR-100

Features

Wide variety of sample tube applications

・Accommodates angle rotor or swing out rotor
・Can load a variety of sample tubes, such as 15 ml, 50 ml, 250 ml*
・Max. speed at 10,000 rpm for 50 ml sample tubes**
*Swing out rotor TS-33R
**Angle rotor CA-16R

Accessories

Swing rotor
TS-33R[50 ml~250 ml]
Angle Rotor
CA-16R[50 ml conical tube ×8]

Specification

Product Bench-Top Step Stop Centrifuge
Model LCXR-100
Rotor Stopping potions 4 positions (Rotate Angle every 90°)
Max. Speed 10,000 rpm
Max. RCF 11,740 G (Angle rotor CA-16R)
Max. Capacity 1,000 ml (Swing out rotor TS-33R)
Speed Setting Range 100 rpm to 10,000 rpm (100 rpm increments)
Time Setting Range From 10 sec to 50 sec, in 10 sec increments,
From 1 min to 99 min, in 1 min increments or for free
Dimension 430W × 582D (Lid Open 690D) × 514H (Lid Open 940H) mm
Net Weight Approx 49 kg
Power Requirements 1-Phase AC100 V 50 / 60 Hz 15 A
(Compatible with 115 V)
Rated Current 8 A
Power Consumption 430 W (370 Kcal / h)
External Interface Terminal 1:Contact closure signal input (Select the rotor defined positions from 1 to 4, Start)
Output during power-on
Terminal 2:PATLITE Corporation's LED Signal Tower Lights
Safety Devices Imbalance Detector
Abnormal Speed Detector
Lid Interlock
Overcurrent Circuit Breaker(Rated breaking Current 15 A)
Lid Open / Close Detector
Rotor Identification System
Motor Overcurrent Detector

Product Video

Bench-Top Step Stop Centrifuge LCXR-100

Step Stop Mixer-Centrifuge SMCR-006